PHL-A
    
    
		    	
		        Compact, Lightweight, Usable in Both High and Low Temperatures, and High Frequency Response.
PHL-A- series pressure transducers are suitable for pressure measurement in not only limited space under both high and low temperature environments but also highly viscous heated fluids like melt resin, high-temperature gases and LPG/LNG tanks. A semi-flush diaphragm at the top not only ensures excellent response but also dynamic characteristics. (Patented).
Connector-equipped PHL-A-A is also available.
 
        
    
					 
    
      | Model Name | Rated Capacity | Natural Frequency (Approx.)
 | 
    
      | Connector Type | Cable Integrated Type | 
    
      | PHL-A-1MP-A | PHL-A-1MP-B | 1MPa | 48kHz | 
    
      | PHL-A-2MP-A | PHL-A-2MP-B | 2MPa | 74kHz | 
    
      | ----- | PHL-A-3MP-B | 3MPa | 94kHz | 
    
      | PHL-A-5MP-A | PHL-A-5MP-B | 5MPa | 122kHz | 
    
      | PHL-A-10MP-A | PHL-A-10MP-B | 10MPa | 149kHz | 
    
      | PHL-A-20MP-A | PHL-A-20MP-B | 20MPa | 210kHz | 
    
      | ----- | PHL-A-30MP-B | 30MPa | 250kHz | 
     
      | PHL-A-50MP-A | PHL-A-50MP-B | 50MPa | 294kHz | 
  
	| Performance | 
	| Rated Capacity | See table above | 
	| Nonlinearity | Within??0.5% RO (PHL-A-1 and 2MP) Within ??0.3% RO (PHL-A-3 to 50MP)
 | 
	| Hysteresis | Within??0.5% RO (PHL-A-1 and 2MP) Within ??0.2% RO (PHL-A-3 to 50MP)
 | 
	| Repeatability | 0.2% RO or less | 
	| Rated Output | 2 mV/V (4000??m/m) ??20% | 
	| (??30% with PHL-A-1 and 2MP) | 
	| Environmental Characteristics | 
	| Safe Temperature Range | -40 to 150??C | 
	| Compensated Temperature Range | -196 to 210??C | 
	| Temperature Effect on Zero Balance | Within??0.05% RO/??C(PHL-A-1 and 2MP) | 
    | Within??0.03% RO/??C (PHL-A-3 to 50MP) | 
   	| Temperature Effect on Output | Within??0.03%/??C | 
	| Electrical Characteristics | 
	| Safe Excitation Voltage | 6V AC or DC | 
	| Recommended Excitation Voltage | 1 to 3V AC or DC | 
	| Input Resistance | 350???2% | 
	| Output Resistance | 350???2% | 
	| Cable | 4-conductor (0.08 mm 2 ) fluoroplastic shielded cable, | 
	| 3.1 mm diameter by 30 cm long, terminated with connector plug | 
	| (Shield wire is not connected to mainframe.) | 
	| Mechanical PropertiesConnector | 
	| Safe Overload Rating | 150% | 
	| Natural Frequency | See table above | 
	| Material | SUS 630 (liquid-contacting part) | 
	| Weight | Approx. 20 g (not including cable) | 
	| Mounting Screw | G1/8, male | 
			 
    
    
    
    
    
        
    
					Gasket (mild copper) (SS-105 O-ring is also usable.)